Example: |
POLIMOBIL III Manufacturer Siemens: |
Specifications: |
XRAY GENERATOR and CONTOLS |
Main voltage: |
230V +/-10% 50/60Hz, fuse 10A |
|
110V +/-10% 50/60Hz, fuse 15A |
Internal line impedance: |
Max. 0.9ohms at 230V |
|
Max. 0.3ohms at 110V |
Output: |
2.5kW (according to EN/DIN 6822) |
Wave shape: |
Multi-pulse, max 5kV ripple |
KV range: |
40-100kV in 21 steps |
mA range |
25-40mA |
mAs range (tube current time product) |
0.32-200 mAs in 29 steps |
|
0.5-200mAs at 40kV |
|
0.5-160mAs at 41-48kV |
|
0.4-160mAs at 50kV |
|
0.4-125mAs at 52-57kV |
|
0.32-100mAs at 66-77kV |
|
0.32-125mAs at 60-63kV |
|
0.32-80mAs at 81-100kV |
Shortest exposure time: |
8ms |
Max exposure time: |
5s |
Length of power cord: |
5m, cable holder attached to Polymobil III |
|
|
|
TUBE HOUSING ASSEMBLY |
X-ray tube: |
Siemens SR/100/20 with nominal focal spot value 1.4 IEC 336/1982 |
Application: |
brief operation, according to exposure table. |
Collimator: |
manually adjustable, can be rotated at ±45° |
Light localiser: |
halogen light bulb 12V, 180 lux at 1m FFD. |
Maximum cassette size: |
43x43cm |
Weight: |
153kg |
Max floor incline for transport: |
<10° |
Dimension LWH: |
tube in park position, 1300x700x1325mm |
Tube positioning: |
maximum focal spot height 1915mm |
Minimal focal spot height: |
430mm |
Maximum horizontal extension: |
805mm |
Normal use: |
temperature +10° to 40°C, humidity 30 to 70%, atmospheric pressure 700 to 1060hPa |
Installation: |
should be checked by manufacturer-agreed distributor and a trained technician. |
Warranty: |
12 months from the date of installation or 15 months from the invoice date, whichever comes first. |
Maintenance contract: |
locally via local distributor, every 12 months. |
|
To order spare parts always indicate the serial and reference numbers indicated in the user and maintenance manual. |
|
Expected delivery time from supplier: 4-6 weeks. |
Transport and storage: |
Temperature -20° to + 70°C, humidity 10 to 100%. |
Atmospheric pressure: |
500 to 1060hPa |
Packaging: |
Unit presentation: complete unit with user’s manual and instructions. |
Size LWH |
136x84x155cm |
Total net weight: |
142kg |
Size packed: |
1 box 140x90x160cm |
Total gross weight: |
192kg |
Volume: |
1,795m³ |
|
The following information should appear on the outside packaging: |
|
- Designation of item, multilingual. |
|
- Serial number. |
|
- Name and address of supplier (manufacturer). |
Other: |
Conforms to EN 60601-1 and EN 60529. |
Part of: |