X-ray unit basic, mobile, 40 mA, Polymobil III, Siemens

last updated : 2018/12/31
Code Description Budget Price Ship. weight Ship. vol.
XXRAUNITM4B
BULB, 12V, 100W, Osram, Xenophot-HLX,HLX 64625, FCR on request 0kg 0 Add
XXRAUNITM4F
(X-ray unit Polymobil III) FUSE, spare CH 129.77 0kg 0 Add
XXRAUNITM4
X-RAY UNIT, BASIC, MOBILE, 40 mA, 'Polymobil III', Siemens CH 17,972.77 192kg 1,795248m³ Add
Item to be justified

General information

A basic mobile X-ray unit used in emergency situations and in field hospitals.

The unit disassembles easily; the operating console and the X-ray tube can be removed from the mobile base.

Can be operated from any standard and grounded wall outlet.

Operating personnel need to familiarize themselves with operating instructions prior to use.

Specifications

Example:

POLIMOBIL III Manufacturer Siemens:

Specifications:

XRAY GENERATOR and CONTOLS

Main voltage:

230V +/-10% 50/60Hz, fuse 10A

 

110V +/-10% 50/60Hz, fuse 15A

Internal line impedance:

Max. 0.9ohms at 230V

 

Max. 0.3ohms at 110V

Output:

2.5kW (according to EN/DIN 6822)

Wave shape:

Multi-pulse, max 5kV ripple

KV range:

40-100kV in 21 steps

mA range

25-40mA

mAs range (tube current time product)

0.32-200 mAs in 29 steps

 

0.5-200mAs at 40kV

 

0.5-160mAs at 41-48kV

 

0.4-160mAs at 50kV

 

0.4-125mAs at 52-57kV

 

0.32-100mAs at 66-77kV

 

0.32-125mAs at 60-63kV

 

0.32-80mAs at 81-100kV

Shortest exposure time:

8ms

Max exposure time:

5s

Length of power cord:

5m, cable holder attached to Polymobil III

 

 

 

TUBE HOUSING ASSEMBLY

X-ray tube:

Siemens SR/100/20 with nominal focal spot value 1.4 IEC 336/1982

Application:

brief operation, according to exposure table.

Collimator:

manually adjustable, can be rotated at ±45°

Light localiser:

halogen light bulb 12V, 180 lux at 1m FFD.

Maximum cassette size:

43x43cm

Weight:

153kg

Max floor incline for transport:

<10°

Dimension LWH:

tube in park position, 1300x700x1325mm

Tube positioning:

maximum focal spot height 1915mm

Minimal focal spot height:

430mm

Maximum horizontal extension:

805mm

Normal use:

temperature +10° to 40°C, humidity 30 to 70%, atmospheric pressure 700 to 1060hPa

Installation:

should be checked by manufacturer-agreed distributor and a trained technician.

Warranty:

12 months from the date of installation or 15 months from the invoice date, whichever comes first.

Maintenance contract:

locally via local distributor, every 12 months.

 

To order spare parts always indicate the serial and reference numbers indicated in the user and maintenance manual.

 

Expected delivery time from supplier: 4-6 weeks.

Transport and storage:

Temperature -20° to + 70°C, humidity 10 to 100%.

Atmospheric pressure:

500 to 1060hPa

Packaging:

Unit presentation: complete unit with user’s manual and instructions.

Size LWH

136x84x155cm

Total net weight:

142kg

Size packed:

1 box 140x90x160cm

Total gross weight:

192kg

Volume:

1,795m³

 

The following information should appear on the outside packaging:

 

- Designation of item, multilingual.

 

- Serial number.

 

- Name and address of supplier (manufacturer).

Other:

Conforms to EN 60601-1 and EN 60529.

Part of: