Component: |
ULTRA SOUND BASIC UNIT, ALOKA 500, with convex probe 3.5Mhz. |
|
1 scanner, ultrasound portable system, main unit with accessories. |
|
1 standard electronic convex sector probe: frequency 3.5MHz, scanning angle 60°, radius 60mm. |
|
1 ultrasonic gel 5-litre container. |
|
1 dispenser for ultrasonic gel. |
|
1 operating manual: English and French |
|
1 service manual: English and French |
Specifications: |
Portable high-performance ultrasound system, main unit. |
|
Compact and lightweight, easy to carry. |
|
Size WHD 29x32x25cm, weight: 10kg |
|
Monitor with screen: simple controls, straight-forward operation. |
|
Scanning method: electronic convex sector probe, electronic linear. |
|
Imaging display modes: B, dual B, M, and B and M. |
|
Imaging technologies: dynamic frequency imaging, multi-stage focusing and aperture control. |
|
Image gray-scale: 64 levels. |
|
Depth range selection: convex sector image and linear image, 3 steps. |
|
Image orientation: lateral and vertical inversion (in B mode). |
Measurement and analysis: |
Calibre control: trackball. |
|
On B-mode image: distance, area and circumference via ellipse or trace method; and volume, ratio, gestational age, foetal weight, angle. |
|
Gestational table: user programmable. |
|
On M-mode: velocity, time interval, depth, heart rate, LV function. |
Characters & graphics: |
Patient information, annotation area, body mark display. |
|
Automatic display: date and time, focal-point setting, image orientation mark, image scrolled position, distance scale mark, M-mode time mark, and grey scale for calibration. |
|
Selectable On/Off automatic display: gain setting, near gain setting, far gain setting, probe frequency, and puncture guide line. |
|
Viewing monitor: 7-inch diagonal monitor with non-reflexion filter. |
|
Video output: 625 lines per frame. |
|
Power requirement: 220/240V ±10%, 50/60Hz, approximately 160VA. |
Environmental requirements: |
In operation: temperature -10 to +40°C, relative humidity 30 to 85% non-condensing. |
|
In storage: temperature -10 to +50°C, relative humidity 10 to 90% non-condensing. |
Packaging and labelling: |
Primary packaging: unit of use |
|
Scanner ultrasound, individually wrapped in a plastic film with manufacturer’s instruction for use, spare parts and accessories (including list of). Supplied with clear instructions for use and diagrams for assembly. Must be multilingual: English, French and Spanish, others when available. |
Other requirement: |
Conforms to Medical Device Directive, Council Directive 93/42/EEC and has CE mark and reference number of notifying body. |
Safety regulation: |
Complies with IEC 601-1 Class1, type BF |