Components: |
Sphygmomanometer parts: |
|
- Aneroid pressure gauge (with no stop pin), with air pressure release-valve. |
|
- Rubber bulb with palm fitting handle. |
|
- Connector to rubber tube. |
|
- Connecting tube. |
|
- Cuff with inflatable bladder. |
Materials: |
Hardened copper-beryllium: membrane or diaphragm |
|
Stainless steel: handle |
|
Nickel-plated: tube connector and valve |
|
Nylon: cuff and velcro |
|
Latex: connecting tube, bulb, inflatable bag |
Specifications: |
A sphygmomanometer consists of an inflatable cuff, a measuring unit (the manometer), a tube to connect the two, and an inflation bulb, which is also connected by a tube to the cuff. The inflation bulb contains a one-way valve, to prevent inadvertent pressure leakage, with an adjustable screw valve so that the operator can lower the pressure in the system in a controlled manner. |
|
Manometer, aneroid gauge: |
|
An aneroid manometer is a mechanical gauge with a round dial and a needle that rotates to indicate pressure in mmHg. On the inside, a bellow and spring mechanism senses and transmits the blood pressure measurement to the dial. |
|
Diameter: 49mm |
|
Linear scale: graduated from 0-300mmHg, easy-to-read, maximum error throughout range ±3mmHg. |
|
Diaphragm membrane: can withstand overpressure of up to 600mmHg. |
|
- Connector to rubber tube: on the right-hand side, below the manometer. |
|
- Inflation bulb: for inflating and deflating the rubber-bag cuff. |
|
Mounted with a one-way valve, to prevent inadvertent pressure leakage, with an adjustable screw valve so that the operator can lower the pressure in the system in a controlled manner. |
|
- Cuff: Washable at 30°C, made of non-deformable material with velcro fastening to enable a tight adjustment around the arm. Inside the cuff is an inflatable airbag (bladder). |
|
The airbag (bladder) must be airtight. |
|
3 interchangeable cuff sizes: |
|
XMEQSPHYRIBC05: Length 15.5cm, width 5cm - newborn |
|
XMEQSPHYRIBC07: Length 23cm, width 7.2cm – baby or infant |
|
XMEQSPHYRIBC15: Length 35.5cm, width 10cm - child |
|
- Connecting tube: length 60cm from cuff to aneroid manometer. |
Packaging: |
Unit presentation: in a vinyl or leatherette zippered case. |
|
The following information must appear on the outside packaging: |
|
- Designation of item, multilingual if possible (English, French and Spanish, others when available.) |
|
- Name and address of supplier (manufacturer) and CE mark and reference number of notifying body. |
|
- Enclosed inside the packaging: manufacturer’s certificate of guarantee and multilingual instructions for use and maintenance: English, French and Spanish, others when available. |
Other requirements: |
Conforms to EN 1060-1-1A –2002 and EN 1060-2-1995 |