 Components:  | 
 - Sphygmomanometer parts:  | 
| 
    | 
 Aneroid pressure gauge (no stop pin) with an air pressure release-valve.  | 
| 
    | 
 Rubber bulb with palm fitting handle.  | 
| 
    | 
 Connector to rubber tube.  | 
| 
    | 
 Connecting tube.  | 
| 
    | 
 Cuff with inflatable bladder.  | 
| 
 Materials:  | 
 Hardened copper-beryllium: membrane or diaphragm.  | 
| 
    | 
 Stainless steel: handle  | 
| 
    | 
 Nickel-plated: tube connector and valve.  | 
| 
    | 
 Nylon: cuff with velcro fastener  | 
| 
    | 
 Latex: connecting tube, bulb, inflatable bag.  | 
| 
 Specifications:  | 
 A sphygmomanometer consists of an inflatable cuff, a measuring unit (the manometer), a tube to connect the two, and an inflation bulb, which is also connected by a tube to the cuff. The inflation bulb contains a one-way valve, to prevent inadvertent pressure leakage, with an adjustable screw valve with which the operator can lower the pressure in the system in a controlled manner.  | 
| 
    | 
 Adult, palm style, sphygmomanometer with one tube.  | 
| 
    | 
 Manometer, aneroid gauge:  | 
| 
    | 
 - An aneroid manometer is a mechanical gauge with a round dial and a needle that rotates to indicate pressure in mmHg. On the inside, a bellow and spring mechanism senses and transmits the blood pressure measurement to the dial.  | 
| 
    | 
 Diameter: 63mm  | 
| 
    | 
 Linear scale: graduated from 0-300mmHg, easy-to-read, maximum error throughout range ±3mmHg.  | 
| 
    | 
 Diaphragm membrane: can withstand overpressures of up to 600mmHg.  | 
| 
    | 
 - Connector to rubber tube: on the right-hand side, below the manometer.  | 
| 
    | 
 - Inflation bulb: to inflate or deflate the rubber-bag cuff.  | 
| 
    | 
 Mounted with a one-way valve, to prevent inadvertent pressure leakage, with an adjustable screw valve with which the operator can lower the pressure in the system in a controlled manner.  | 
| 
    | 
 Cuff: Washable at 30°C, made of non-deformable material with velcro fastening to enable a tight adjustment around the arm.   | 
| 
    | 
 Inside the cuff is an inflatable airbag (bladder); it must be airtight. Size: 22x10cm.  | 
| 
    | 
 XMEQSPHYRIAC14: length 54.5cm, width 14cm  | 
| 
 Packaging:  | 
 The following information must appear on the outside packaging:  | 
| 
    | 
 - Designation of item, multilingual if possible (English, French and Spanish, others when available.)  | 
| 
    | 
 - Name and address of supplier (manufacturer) and CE mark and reference number of notifying body.  | 
| 
    | 
 - Enclosed inside packaging: manufacturer’s certificate of guarantee and multilingual instructions for use and maintenance: Must be multilingual: English, French and Spanish, others when available.  | 
| 
 Other requirements:  | 
 Conforms to EN 1060-1-1A –2002 and EN 1060-2-1995  |