Components: |
Ambidextrous powdered glove composed of 5 fingers, a palm and a sleeve. |
Materials: |
Natural latex |
|
Powder: resorbable maize starch |
Specifications: |
Ambidextrous: fits either hand |
|
Meets European norm EN 455-1 |
|
Watertight |
|
Meets European norm EN 455-2 |
|
Moulded manufacture |
|
Total length: 240mm minimum |
|
Width: |
|
- Small: 80 ± 10mm |
|
- Medium: 95 ± 10mm |
|
- Large:110 ± 10mm |
|
Strength: breaking force (for latex gloves): |
|
- Before accelerated ageing: 7.5N |
|
- After accelerated ageing: 5.5N |
|
Low residual protein level (refer to EN 455-3): |
|
- 250µg/g for powdered gloves |
|
- 50µg/g for non-powdered gloves |
|
Lengthening at break: minimum 600% |
|
Cuff length: minimum 5cm |
|
Strictly for single use. |
|
After use, ensure complete destruction by incinerating in a protected area. |
Storage: |
In a dry place, away from a direct heat source. |
Packaging: |
Cardboard dispensing box with 100 gloves. |
|
Labelling on the box: Manufacturer’s name and address; product name and size; manufacturing date and expiry date; and CE mark. Must be multilingual: English, French and Spanish, others when available. |
|
Complies with European Norm EN 455-3: |
|
Gloves that are directly derived from latex need to indicate the following: "(product) containing natural rubber latex likely to provoke allergic reactions". |
|
- Powdered gloves need to bear a label specifying that they are powdered |
Other requirements: |
Must comply with European Medical Device Directive 92/42/EEC, or equivalent international norm. |
Meet standard: |
EN 455-1:2000: Medical gloves for single use – Part 1: Requirements and testing for freedom from holes |
|
EN 455-2:2000: Part 2. Requirements and testing for physical properties |
|
EN 455-3:2000: Part 3. Requirements and testing for biological evaluation |